Complex Oxygen Transducer and Experimental Research for Oxygen Partial Pressure on Acupoint |
Hits 1527 Download times 1784 Received:August 20, 2005 |
View Full Text View/Add Comment Download reader |
DOI
10.11656/j.issn.1672-1519.2005.06.32 |
Key Words
Complex transducer;needle-type;partial pressure of oxygen;acupoint |
Author Name | Affiliation | YIN Fei | Tianjin University, Tianjin 300072, China | GAO Jian-ping | Tianjin University, Tianjin 300072, China | GUO Yi | 天津中医学院, 天津, 300193 | 苗文方 | 天津中医学院, 天津, 300193 | 李建宇 | 天津中医学院, 天津, 300193 | 王卫 | 天津中医学院, 天津, 300193 | 魏连海 | 天津中医学院, 天津, 300193 |
|
Abstract
|
[Objective] To work out a transducer of needle-type for the measurement of the oxygen partial pressure in vivo.[Methods] The indicatingr electrode was made with the inert metal as the sensitive membrane. It was installed in a syringe needle with a reference electrode together to establish a complex needle-type transducer. It was in solid state with the diameter of needle tip of 0.35 mm,sensitivity 0.1 kPa,and linearity range 0.1~20 kPa. The detector of oxygen partial pressure with four channels was then manufactured with the polarization voltage -0.4~-0.9 v,measurement range 0~20 kPa,sensitivity 0.01 kPa.[Results] The experiment in vitro and in vivo demonstrated that the transducer could measure the oxygen partial pressure of tissue in vivo continually and dynamically and show the changes accurately on time.[Conclusion] The complex oxygen transducer can monitor the oxygen partial pressure of tissue in vivo continually. |
|
|
|
|
|
|